Product Description
HY0871 Ignition temperature tester is used to test the combustion and flame retardant properties of plastic materials. In the heating chamber of a hot air furnace, the samples are heated at different temperatures and the escaping gas is ignited directly by a small flame in the opening above the furnace to determine the flashover temperature. The auto-ignition temperature is determined in the same way as the flash-over temperature, but no flame is applied.
Standards
ISO 871 | GB 4610 |
Technical Specification
Model | HY0871 |
Furnace tube | inner diameter 100mm±5mm, length 240mm±20mm; Made of custom ceramic tubes with a temperature resistance of not less than 750°C |
Inner ceramic tube | Capable of withstanding at least 750℃, inner diameter 75mm±2mm, length 240mm±20mm |
The temperature in the furnace | Can be constant at any temperature between (150-750) ℃ |
Time range | 0~999.9S, settable |
Flame height | 10mm to 20mm, adjustable |
Thermocouples | 3 thermocouples and temperature meters to detect the temperature in the furnace and the temperature of the sample |
Temperature control accuracy | ±2℃ |
Display accuracy | ±0.5℃ |
Test furnace size | 400×250×300mm |
Controller size | 430×350×280mm |
Power supply | AC220V±10%, 50Hz |
Power | 2500W |
Weight | 15kg |