Availability: | |
---|---|
Product Description
AT902 Automated Filter Tester is a universal tester, development according to the standard of NIOSH 42 CFR 84, EN 149:2009, it mainly used for detecting the efficiency of the particulate respirator filtration and detecting the filtration efficiency and resistance of material. Using high precision photometer as detector to ensure accurate results.
The tester is divided into two categories, oil and salt fog. The oil spray tester is using DEHS as aerosol and the salt spray tester using NaCl, the solution spray after Ruskin nozzle and drying, then form in to aerosol. The system adopts precision instrument to control flow and concentration, and through integrate of system to communicate with computer, the system has good stability, and it is easy to operate.
By detecting the concentration of upstream and downstream, and then calculate the results through the formula. Airflow can be real-time monitored through electric proportional valve and flow meter, to ensure the accuracy of the test flow and the stability. The aerosol concentration can be monitored by electrical proportional valve and differential gauge, and provide stable test particles.
The filtering material tester is a fully automatic instrument which can control the following actions by a computer:
▲It can feedback control and read all of the working parameters automatically: including the volume, the resistance, test time and the efficiency.
▲ Measuring the particle number and the resistance of sample in a test period.
▲Showing operating parameters.
▲The system makes use of serial communication to initiating a transfer of the data files
Standards
EN143 | GB2626-200 | JMOL |
NIOSH 42 CFR 84 |
Machine Structure
Features
Structure
1. Split-type design, customer can change the generator easily.
2. Aluminium profile frame, nice and easy to clear.
3. The machine can be calibrated and all the origin date is available.
Multifunction
1. This machine can be used for different filter material
2. Customized clamp for different sample size is available
3. Both salt solution and oil solution test can run on this machine
High Accuracy
1. Adopted high quality photometer
2. Equipped with oil water separator to avoid water effect test result
3. Test date accuracy reach to 0.00001%
Software
1. Support Chinese and English, menu type operating mode and large LCD display, it’s easy to switch
2. Two test method available: efficient test and load test
3. Simple, fast and automated operation
Technical Specification
Model | AT902 |
Detector | photometer |
Air volume | (20〜100) L/min |
Efficiency | 10%-99.99999% |
Test time | ≤14S |
Loading efficiency | 10%-99.999% |
Photometer detect aerosol concentration range | 0.001~200 mg/m3 |
Aerosol | Salt, DOP, DEHS, Paroline, Corn oil, optional |
Resistance | 0-2000Pa |
Aerosol density | (50〜200) mg/m3(oil); (12〜20)mg/m3(salt) |
Detector | photometer |
Measuring range | 0.3~1μm |
Counting median diameter | 0.075±0.02μm (salt) |
0.185±0.02μm(oil) | |
Detect temperature | 5~ 35℃ |
Related humidity | 0 ~ 90% RH |
Pressure detect | Digital pressure sensor (pressure difference transmitter) |
Pressure detect range | (0~0.88)MPa |
Efficiency testing range | filter efficiency: 10~99.99999% |
Operation display | Computer |
Power supply | 110V/220V, 50/60Hz |